Difference between revisions of "Using the echoMRI"
From Bridges Lab Protocols
Davebridges (Talk | contribs) (wrote initial page) |
(No difference)
|
Revision as of 13:06, 15 November 2016
Before you start
- Make sure the instrument and computer are turned on
- Open the software if its not already open
- Create a new folder for your experiment in the format YYMMDD-NAMEOFSTUDY
- Make sure the mouse antenna is connected if you are doing mice, or its taken off if you are doing rats
- Change the selector to Mice or Rats as appropriate
- Change the water to No (unless you really need free water determination)
- Place the canola oil tube in and click System Test
Each Sample
- Weigh the animal and record the weight.
- Gently place the mouse/rat in the appropriate sized tube and using the plunger compress it at the end. The animal should be snug but not so tight that it is hurt. If the animal is moving then he is probably not snug enough.
- Select Start Measurement or hit F5 and enter the mouse ID, weight and any comments then hit run
- If you have a lot of animals, then load up the next tube while the first one is being run.
- If you mis-label an animal then click on Edit Comments
Exporting Data
- Select all your measurements (CTRL-A or shift and point)
- Go to export table and export it. Change the filename if needed
- Copy the entire folder to a USB stick for analysis