Triglyceride Assay from Cells and Tissues
From Bridges Lab Protocols
Revision as of 16:03, 13 April 2012 by Davebridges (Talk | contribs) (added details about using the sigma kit)
Materials
- Homogenization Buffer (50 mM Tris, 5 mM EDTA, 30 mM Mannitol, PI inhibitor)
- 10M KOH
- Chloroform/Methanol Mixture (2:1)
- Butanol Mixture: 3 mL butanol, 1.66 mL Triton-X114, 0.33 mL Methanol
- Sigma Triglyceride Assay Kit (Cat 337-B)
Protocol
- Weigh out 200-500mg tissue (record weight for normalization)
- Homogenize with dounce homogenizer in 2 mL Homogenization Buffer.
- Remove 200 uL to a tube containing 5 uL KOH
- Mix by inverting
- Add 800 uL Chloroform/Methanol Mixture
- Vortex vigorously then sit at room temperature for 5 min
- Centrifuge 10min at 13 000 RPM
- Take 180 uL of the bottom layer into a fresh tube.
- Dry in fume hood overnight (or until completely dry)
- If absorbance is going to be measured by cuvette, use non-bolded values. If you are using a plate reader use bolded values.
- Add 50uL (500uL) of Butanol Mixture
- Measure triglyceride levels using Sigma Diagnostic Kit using 5 uL of sample:
- Resuspend triglyceride and glycerol reagent with water if necessary.
- Calculate how many samples you have (samples + buffer blank + 6 standard curve values).
- Prepare reagent, you need 560 uL (80uL) of glycerol reagent and 140 uL (20uL) of triglyceride reagent. Make a bit extra and combine in a falcon tube.
- Aliquot 700 uL into a cuvette or 100 uL into a well of a 96 well plate.
- For standards add 0-5 uL of glycerol standard (or of a 1/10 dilution of the glycerol standard).
- Add 5 uL of resuspended lipid to each well to start (also make a 5uL blank of the butanol mixture) and mix.
- Let sit for ~30 min at room temperature (or 5 min at 37C if you are in a hurry).
- Measure absorbance at 540 nm.
- If any samples are A540<0.1 or above the 5uL standard A540 then repeat with more or less lipid as required.