Measuring Grip Strength
From Bridges Lab Protocols
Materials
- Grip Strength Meter: Use the Metek DFE Grip Meter from the Neuroscience Institute Behavioral Core
- Mice: You can either do it downstairs or bring the mice upstairs. If you move the mice then make sure they have some time to settle first.
- Prepare a data sheet with one row per animal and space for 3-5 measurements per animal.
Protocol
- Make sure the meter is charged or plugged in and sitting stably on the bench
- Turn it on, click on the top left button (MODE) until the measurement display says T-PK (Peak Tension) not NORMAL or C-PK (Compression Peak)
- Zero the reading by pressing ZERO, the reading should be at or near 0.00 N (measure force in Newtons unless a compelling reason to do this otherwise)
- Pick up the mouse from the cage, and determine its ear tag number.
- Place the mice on the grate so that is securely grips with all four paws, holding it by the tail.
- Gently pull the mouse horizontally by the tail away from the sensor until the mouse lets go.
- Record the peak grip on your data sheet
- Press zero to reset the instrument
- Let the mouse recover for 2-3s, then repeat 3-5 times in total
- Move on to the next mouse